SDM4075A — 7½ digits, 16 ppm, 50,000 readings per second
The SDM4075A is the top of Siglent’s bench multimeter range: a 7½-digit instrument with 11,000,000 counts, 16 ppm basic DC-voltage accuracy over a year and the low noise and linearity that calibration and reference work demand, built on the same 5-inch touchscreen platform as the SDM4055A and SDM4065A.
It is also fast: 50,000 readings per second at the shortest integration time, a digitize mode sampling at up to 2 MSa/s with 15 kHz bandwidth, and a 2,000,000-reading memory with time-stamped logging for up to 100 hours. Front and rear terminal sets are selected by a mechanical switch, so a permanent fixture at the back and a probe at the front share one meter. DC and AC voltage to 1000 V, current to 10 A (30 A with the SCD30A shunt), 2- and 4-wire resistance to 1 GΩ, capacitance, frequency and RTD/thermocouple temperature complete the function set, with SCPI over USB and LAN, VNC and a web server for automation.
Key features
- Real 7½ digits: 11,000,000 counts, 16 ppm basic DCV accuracy (1 year)
- 50,000 readings per second at 0.001 PLC; 7½ digits at 10 and 100 PLC
- Digitize mode: 10 Sa/s to 2 MSa/s, 15 kHz bandwidth, 2,000,000 readings per capture
- Data log 1 ms to 3600 s intervals: 2 million readings in memory, 360 million to file, 100 hours
- DCV 100 mV–1000 V, ACV to 750 V True-RMS, DCI 1 µA–10 A, ACI to 10 A; 30 A with the SCD30A shunt
- 2/4-wire resistance 10 Ω–1 GΩ, capacitance 1 nF–100 mF, frequency 3 Hz–1 MHz, RTD, thermocouple and user sensors
- >10 GΩ input on the 100 mV–10 V ranges, 50 pA bias current, 140 dB CMRR, autozero
- Front and rear terminals with a mechanical selector; 5-inch 800 × 480 touch display with four views, dual display and probe hold
- Auto, single, external and level trigger; external trigger input and VMC output on BNC
- USB device, USB host and LAN with SCPI, VNC and web server; GPIB via the USB-GPIB adapter; EasyDMM-X software
Application examples
Calibration and reference work
Sixteen ppm of basic DC accuracy, 7½ digits and autozero put the SDM4075A in transfer-standard territory: compare Zener references, qualify sources and verify that a calibration-lab result reproduces on your own bench, with the >10 GΩ input leaving high-impedance references unloaded.
Transient and inrush capture
Digitize at up to 2 MSa/s with 15 kHz bandwidth into 2 million readings: inrush current profiles, soft-start ramps and relay bounce appear as a waveform record, with level or external triggering and pre-trigger samples to catch the event.
Automated production and long-term test
50,000 readings per second over SCPI, four trigger modes, a VMC handshake output and 100-hour logging make it a drop-in for ATE racks and soak tests; rear terminals keep the fixture wiring permanent while the front stays free for bench work.
Low-level and component measurement
Four-wire resistance from 10 Ω full scale resolves milliohm contacts and shunts; 1 µA and 100 mV ranges with 50 pA bias current read sensor leakage and offsets that coarser meters bury in their own noise.
Precision, speed and a touchscreen to see both

Four views on a 5″ multi-touch display
Numerical, bar meter, trend chart and histogram, plus a dual display for a second parameter and probe hold that freezes the reading when you lift the leads. Statistics, limits, dB/dBm and DCV ratio run on the meter itself.

Digitize: 2 MSa/s into 2 million readings
The digitize mode samples DC voltage or current at up to 2 MSa/s with 15 kHz bandwidth, pre- or post-trigger, from level, external or internal triggers. Inrush, soft-start and relay bounce become a waveform record, not an average.






Around the instrument
A second set of input terminals on the rear panel, selected by a mechanical switch on the front, together with the external trigger input, VMC output, LAN and USB connectors.




Who it is for and why
Choose the SDM4075A when the SDM4065A’s 6½ digits stop being enough: calibration laboratories and metrology benches comparing references, R&D groups characterising ADCs, regulators and sensors at the microvolt level, and automated test systems that need both the accuracy and the 50,000-readings-per-second throughput. The front/rear terminal switch is the practical difference on a shared bench, and the 2 MSa/s digitizer means one instrument covers precision DC work and transient capture.
- Calibration and metrology laboratories
- Precision R&D: references, ADC/DAC and sensor characterisation
- Automated production test over SCPI, LAN and GPIB
- Long-term reliability and drift monitoring
- Reference and source verification to 16 ppm
- Digitized capture of inrush and power-up events at 2 MSa/s
- Four-wire milliohm and high-resistance (to 1 GΩ) measurement
- 100-hour logged tests with 2 million time-stamped readings
Specifications
| Resolution | 7½ digits · 11,000,000 counts |
| DCV basic accuracy | 16 ppm (1 year, 10 V range) · 8 ppm in 24 hours |
| Reading rate | 50,000 rdgs/s at 0.001 PLC · 5,000 at 0.01 PLC · 500 at 0.1 PLC · 50 (60) at 1 PLC · 0.5 (0.6) at 100 PLC |
| Digitize | DC voltage and current · 10 Sa/s to 2 MSa/s · 15 kHz (DCV) / 10 kHz (DCI) bandwidth · 2,000,000 readings |
| Data logging | 1 ms to 3600 s interval · 2,000,000 readings to memory, 360,000,000 to file · up to 100 hours |
| DC voltage | 100 mV to 1000 V · 10 MΩ or >10 GΩ input (100 mV–10 V) · 50 pA bias · 140 dB CMRR |
| AC voltage (True-RMS) | 100 mV to 750 V · 3 Hz to 300 kHz · 1 MΩ ∥ <150 pF · crest factor ≤5 |
| DC / AC current | DCI 1 µA to 10 A · ACI 100 µA to 10 A, 3 Hz to 10 kHz · automatic 3 A / 10 A range switching · 30 A with the SCD30A external shunt |
| Resistance | 10 Ω to 1 GΩ, 2- and 4-wire · open-circuit voltage <10 V |
| Capacitance | 1 nF to 100 mF |
| Frequency / period | 3 Hz to 1 MHz, reciprocal counting, gate 1 ms to 1 s |
| Temperature | RTD (α = 0.00385), thermocouples B, E, J, K, N, R, S, T, user-defined sensors; built-in or external reference junction |
| Continuity / diode | 1 mA source · continuity threshold 1 Ω to 1 kΩ · diode threshold 0 to 5 V |
| Display | 5-inch 800 × 480 colour touchscreen · numerical, bar meter, trend, histogram · dual display · probe hold · statistics, limits, dB/dBm, ratio |
| Trigger | Auto, single, external, level · delay 0 to 1000 s · 1 to 599,999,999 samples per trigger · external trigger in (5 V TTL) · VMC out |
| Terminals | Front and rear input terminals, selected by a mechanical switch |
| Interfaces | USB device (USBTMC), USB host, LAN (SCPI, VNC, web server) · GPIB via optional USB-GPIB adapter · EasyDMM-X PC software |
| Power | 100–120 V / 200–240 V, 45–66 Hz · 30 W max. |
| Size / weight | 420 × 260 × 103 mm · 4.5 kg |
| Environment | Full accuracy 0 to 50 °C · 90 minutes warm-up · altitude to 2000 m |
| Warranty | 3 years |
From the SDM4075A Data Sheet EN01B. Accuracies apply after 90 minutes of warm-up at 100 PLC; errors and omissions excepted.
In the box
Everything included as standard with the SDM4075A. EasyDMM-X is a free download.






Accessories and related
The SCD30A external shunt extends current measurement to 30 A; the USB-GPIB adapter adds IEEE-488. Need 6½ digits with a 16-channel scanner instead? See the SDM4065A-SC.
External current shunt for measurements beyond the 10 A input, up to 30 A.

Add GPIB (IEEE-488) control for ATE integration.

Same touchscreen platform and 50,000 rdgs/s, 2,200,000 counts, 35 ppm; also as -SC with a 16-channel scanner.
7½ digits for the bench or the rack