Siglent SSA5000-NF Noise Figure Measurement Option for SSA5000A (software licence)

€2,000.00 FREE shipping

SSA5000-NF: Y-factor noise-figure and gain measurement on the Siglent SSA5000A. Amplifiers and converters, ENR table, calibration. Needs the NSD28 driver.

Home › SSA5000A options › SSA5000-NF
SSA5000A software option · noise-figure measurement

SSA5000-NF — noise figure and gain, measured on the spectrum analyser you already have

Software licence — activated by key, no hardware

Noise figure is the number that sets a receiver’s sensitivity: how much a stage degrades the signal-to-noise ratio passing through it. The SSA5000-NF licence gives the SSA5083A and SSA5085A a dedicated noise-figure mode using the Y-factor method: it switches a calibrated noise source between hot and cold at the input of the device, compares the two output powers, applies the source’s ENR table, removes the analyser’s own noise by calibration and shows noise figure and gain against frequency.

The mode measures amplifiers, up-converters and down-converters, reports seven results per frequency point, compensates input and output line loss and displays as graph, table or meter. It needs two more things: the NSD28 driver, which supplies the switched +28 V the noise source expects, and a noise source with its ENR table, which Siglent does not sell.

Y-factor NF and gain Amplifiers and converters Calibration, ENR table, loss compensation SSA5083A and SSA5085A
Siglent SSA5000A noise figure and gain traces of an amplifier from 10 MHz to 10 GHz
Noise figure (top) and gain (bottom) of an amplifier from 10 MHz to 10 GHz in 11 points, noise source drive plugged in and calibration applied.

What the licence adds to the analyser

Calibration with second-stage correction

Measure the noise source directly first; the analyser stores its own noise figure and removes it from every later result. A calibrated system without a device reads 0 dB.

ENR table

Enter the excess noise ratio of your noise source against frequency once, from its calibration sheet, and save it with the set-up.

Seven results per point

Noise figure, gain, noise factor, effective temperature Te, Y-factor, Phot and Pcold, as traces, a table or a meter.

Amplifiers and converters

Device types: amplifier, up-converter and down-converter, so a mixer stage or a receiver front end can be measured, not only a gain block.

Loss compensation

Input and output line loss entered separately, so the cables and adaptors between source, device and analyser do not end up in the figure.

Set-up on the touchscreen

Start, stop and number of points, RBW from 1 Hz to 3 MHz, averaging. Entering the mode switches the preamplifier on and fixes the attenuator at 0 dB.

How the measurement works

Two connections. During calibration the noise source feeds the analyser directly, so the analyser learns its own noise figure. During the measurement the device goes in between. In both the analyser toggles the source through the NSD28, reads the noise power hot and cold, forms the Y-factor and solves for the noise figure of what is in front of it, with its own contribution subtracted.

1 Calibration: analyser + driver + noise sourceSSA5000Anoise-figure modeRF INUSBNSD28driverNoise source+28 V in, RF outRF, directlyUSB control+28 V on / off2 Measurement: the device goes in betweenSSA5000Anoise-figure modeRF INUSBNSD28driverNoise source+28 V in, RF outDUTamplifierUSB control+28 V on / off

Hot and cold

With +28 V applied the source is hot and emits a known excess noise; with the drive off it is cold, a 50 Ω termination at room temperature. The analyser measures the output noise power in both states.

ENR and the Y-factor

Y is the ratio of the hot and cold powers. With the excess noise ratio of the source at that frequency, NF = ENR − 10·log10(Y − 1). Gain follows from the same two readings once the calibration is known.

Second-stage correction

The calibration measures the analyser’s own noise figure. With the device in line that contribution, divided by the device gain, is removed, so the result describes the device alone. That is also why low-gain devices are the hardest case.

Three parts, one measurement

Order the licence and the driver together; choose the noise source for the frequency range and ENR your devices need.

1 · This product
SSA5000-NF licence

The noise-figure mode: calibration, ENR table, averaging, loss compensation, seven results, graph, table and meter. Key issued against the serial number, nothing shipped.

You are hereAdd to your order ↑
2 · Hardware, required
NSD28 noise source driver

USB-A on the analyser side, switched +28 V on a BNC on the noise source side; the mode toggles it during calibration and measurement. Without it nothing switches the source.

€120.00View
3 · Not included
A +28 V noise source

Any source switched by +28 V on a BNC input and supplied with a calibrated ENR table; Siglent names the Keysight 346 family. Typical ENR 6 to 15 dB. Pick the frequency range to cover your devices.

Before you order

The mode switches the analyser’s preamplifier on; Siglent lists the preamplifier as a separate licence (P3 for the SSA5083A, P5 for the SSA5085A), so check your configuration with us if you are not sure it is present. Siglent publishes no measurement uncertainty for noise figure: the ENR uncertainty of your noise source and the device’s gain decide it. Software licences are excluded from the 14-day return right once the key has been issued.

The measurement, step by step

The sequence from Siglent’s application note, measuring a DC–10 GHz low-noise amplifier specified at 26 dB gain and 6 dB noise figure. Result: 26.69 dB and 6.12 dB.

  1. 1
    Warm up and open the mode

    After 30 minutes of warm-up, add a Noise Figure window from the window manager. The status line shows the noise source drive as plugged in once the NSD28 is connected.

  2. 2
    Connect for calibration

    NSD28 into a USB port of the analyser, its BNC to the drive input of the noise source, noise source output straight into the RF input.

  3. 3
    Amplitude

    The mode switches the internal preamplifier on and fixes the input attenuation at 0 dB in automatic mode.

  4. 4
    Frequency

    Set start, stop and points. Siglent’s example: 10 MHz to 10 GHz in 11 points for a DC–10 GHz amplifier.

  5. 5
    ENR table

    Meas Setup → ENR → Edit ENR: frequency and ENR pairs from the noise source’s calibration sheet; save the table.

  6. 6
    Averaging

    Meas Setup → Setting: averaging on, 10 averages in the example.

  7. 7
    Calibrate

    Meas Setup → Cal → Calibrate now. The analyser measures its own noise figure hot and cold and stores the correction.

  8. 8
    Check the zero

    In the table layout a calibrated system reads close to 0 dB noise figure and 0 dB gain at every point.

  9. 9
    Insert the device and read

    Noise source output to the DUT input, DUT output to the RF input. Noise figure and gain appear per frequency, as traces or a table.

On the bench and on the screen

Connection photographs and screens from the application note, in the order of the procedure.

Calibration set-up, from the Siglent application note
Calibration connection: the noise source feeds the RF input directly; the NSD28 on a USB port drives it.
Measurement set-up, from the Siglent application note
Measurement connection: the amplifier under test between the noise source and the RF input.
ENR Edit dialog in noise-figure mode
The ENR table: frequency and excess noise ratio pairs from the noise source’s calibration data, 5.51 dB at 10 MHz to 5.01 dB at 8 GHz in this example.
Calibrated system, graph layout
After calibration, without a device: noise figure and gain at 0 dB across 10 MHz to 10 GHz.
Results in table layout
Table layout: noise figure and gain per frequency point.
Amplifier result, graph layout
The DC–10 GHz low-noise amplifier of the example: noise figure (top) and gain (bottom) in 11 points; 6.12 dB and 26.69 dB against a 6 dB / 26 dB specification.

Where it is used

Low-noise amplifiers and gain blocks

Noise figure and gain across the band, with the analyser’s own noise removed. The classic use, and the one in Siglent’s example.

Receiver front ends and converters

Up- and down-converter device types for mixer stages, LNBs and complete front ends, with the ENR applied at the input frequency.

Cascaded stages

Measure each stage and the chain: the first stage’s noise figure and gain decide the sensitivity of everything behind it.

Production and incoming inspection

A saved ENR table and a calibrated set-up make noise figure a repeatable reading on the analyser already on the bench, no dedicated noise-figure meter.

Design verification

Compare a prototype with its simulation, and watch how bias, matching and temperature move the figure, point by point.

Teaching

Hot and cold states, ENR, Y-factor and second-stage correction, with every intermediate value (Phot, Pcold, Y) available in the table.

Specification

From the SSA5000A data sheet EN02A (noise-figure option table, page 17) and Siglent’s application note.

OptionSSA5000-NF noise-figure measurement, software licence activated by key on the analyser
MethodY-factor: NF = ENR − 10·log10(Y − 1), Y = Phot / Pcold; calibration stores the analyser’s own noise figure as second-stage correction
Supported device typesAmplifier, up-converter, down-converter
ResultsNoise figure, gain, noise factor, effective temperature, Y-factor, Phot, Pcold
Result displayGraph, table, meter
RBW range1 Hz to 3 MHz
Loss compensationInput line loss and output line loss
Analyser settings in the modePreamplifier on, input attenuation 0 dB (automatic); start, stop, points, averaging user-set
Noise source interfaceUSB-A host to BNC (+28 V) via the NSD28 driver
Noise sourceNot included: a +28 V switched source with a calibrated ENR table (Siglent names the Keysight 346 family); typical ENR 6 to 15 dB
Frequency rangeThe overlap of the analyser (9 kHz to 13.6 or 26.5 GHz) and the noise source
AnalysersSSA5083A and SSA5085A
Measurement uncertaintyNot published by Siglent; it depends on the ENR uncertainty of the source, the match and the gain of the device

Questions about SSA5000-NF

What else do I need?

The NSD28 noise source driver, which turns a USB port of the analyser into the switched +28 V the noise source needs, and a noise source with a calibrated ENR table. The licence, the driver and the source are three separate items; the source is not a Siglent product.

Which noise sources work?

Any source switched by +28 V on a BNC input and supplied with an ENR table. Siglent’s support centre names the Keysight 346 family; sources from Noisecom and others built to the same convention behave the same way. Tell us which one you have and we check.

How accurate is the result?

Siglent publishes no uncertainty figure. The dominant terms are the ENR uncertainty of the noise source, mismatch, and the gain of the device: a low-gain device leaves more of the analyser’s own noise in the result even after correction.

Why calibrate first, and how often?

The calibration measures the analyser’s own noise figure with the source connected directly and subtracts it later. Repeat it when the frequency list, the ENR table or the analyser settings change, or after the instrument has warmed up further.

Can it measure mixers and converters?

Yes. Up-converter and down-converter are supported device types next to amplifier. Ask us about the LO arrangement for your converter before you set it up.

Is the preamplifier needed?

Entering the mode switches the preamplifier on. Siglent lists the preamplifier separately as licence P3 / P5 for the two models; if your analyser does not have it, check with us before ordering NF.

How is it activated?

By licence key, issued against the analyser’s serial number and entered in the options menu; nothing is shipped. Software licences are delivered electronically and are excluded from the 14-day return right once the key has been issued.

Is there a cheaper way to estimate noise figure?

For a high-gain, high-NF device Siglent’s application note shows a gain-method estimate from the noise marker and DANL. It is a sanity check with the amplitude accuracy of the analyser; the Y-factor measurement with this option is the one to report.

Noise figure on your SSA5000A

Tell us the devices and frequencies. We confirm the licence, the NSD28 and a suitable noise source, and quote the set as one order.